SC3T Low Helium

Incident Report for CAMRIS

Monitoring

Diagnostic cable to the compressor was knocked loose. Cable is re-attached and alarm is gone. System is fully operational, and users are able to scan. Will continue to monitor throughout the week for any additional leaks or problems.
Posted Jan 06, 2026 - 12:21 EST

Update

Per Siemens engineer: "It is OK to continue scanning, the system is at 50.4% helium and has been steady since ramping yesterday. Temperatures and pressure all look good, except the cold head compressor is not reporting correctly, which is most likely what triggered the alarm. Will be over to assess shortly."
Posted Jan 06, 2026 - 11:50 EST

Update

Ticket#400113613594. Waiting for further instruction from Siemens.
Posted Jan 06, 2026 - 11:33 EST

Investigating

SC3T is showing a low helium warning. CAMRIS staff. Are looking into the issue. In the interim, the scanner is not available for use.
Posted Jan 06, 2026 - 10:55 EST

Monitoring

The SC3T has been ramped up and is ready for use. System is fully operational

Siemens will continue to monitor the secondary cooling loop throughout the week.
Posted Jan 05, 2026 - 16:58 EST

Update

Siemens is currently ramping the SC3T scanner back up to field. We expect to be fully operational by this evening and scanning will resume as scheduled tomorrow.
Posted Jan 05, 2026 - 11:06 EST

Update

Per Siemens service "The current plan is to complete the de-ice and helium fill today, let the system stabilize over the weekend, and ramp up on Monday." Hopeful to have the system back by Tuesday but will provide updates as Siemens works on the issues.
Posted Jan 02, 2026 - 11:21 EST

Identified

The SC3T scanner will now be down through at least Monday January 5th as Siemens completes maintenance, refills helium, and ramps the magnet back to field. We will provide further updates as we have more information.
Posted Dec 31, 2025 - 18:17 EST

Update

SC3T will be offline through at least Friday Jan 2nd as we await a helium refill. Siemens will also perform scheduled service on the magnet during this downtime.

We will provide more information about when the scanner will be available as Siemens investigates the cause of the error and performs service.
Posted Dec 29, 2025 - 14:35 EST

Investigating

Siemens detected a chiller failure at SC3T. The scanner is offline until further notice. We will provide an update on scanner availability when we have more information.
Posted Dec 29, 2025 - 11:09 EST
This incident affects: Stellar Chance 3T (MRI Scanner).